British Journal of Radiology · 1947 · 38 citations · 13 references
Natural SciencesMaterials EngineeringMaterials ScienceRadiation MeasurementRadiation PhysicsSpectroscopyRadiation SpectroscopyX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray OpticsX-ray SpectroscopySpectral CharacterizationSpectral Energy Distribution FittingSpectral Energy Distribution ModelingTarget Material EffectsX-ray Absorption SpectroscopyX-ray DiagnosticsX-ray Spectral AnalysisX-ray Target Materials
1. The Jones-Silberstein method of determining the spectral distribution of the radiation from an X-ray tube is summarised. 2. Its self-consistency is examined experimentally and discussed. 3. Values of absorption data for aluminium, copper and tin resulting from a survey of published material are tabulated, and formulæ are fitted to the values of μ so obtained. 4. The Jones-Silberstein method is extended to cover those cases in which radiation characteristic of the target material is present. 5. The variation with voltage of the ratio of characteristic radiation to continuous radiation is given for the case of tungsten. 6. The accuracy of the method is discussed.
13