Applied Spectroscopy · 1997 · 81 citations · 15 references
Natural SciencesChemistryPharmaceutical AnalysisAnalytical ChemistrySpectroscopic AnalysisSpectroscopyAbsorption SpectroscopyInfrared SpectroscopyInstrumental AnalysisNear-infrared SpectroscopyOptical SpectroscopySpectroscopic Analysis Of PharmaceuticalsDiffuse Reflectance SpectroscopyFiber Optic InstrumentsPartial Least Squares RegressionQuantitative Near-infrared SpectroscopyQuantitative Spectral AnalysisScattering CorrectionSpectral Data PreprocessingSpectral Preprocessing
Near-infrared diffuse reflectance spectroscopy (NIRS) with a fiber-optic probe was used for the determination of the active compound in a commercial pharmaceutical preparation. In order to reduce the strong scatter in the spectra and prevent scatter-induced changes in measurements from prevailing over concentration-induced changes, several data preprocessing methods were tested: normalization, derivatives, multiplicative scatter correction, standard normal variate, and detrending. The effectiveness for reducing the scattering of each data preprocessing was assessed, and the best results were obtained with the use of the second derivative. The effect of the treatments on the quantitation of the active compound by partial least-squares regression (PLSR) was studied, similar results being obtained in all cases, with a relative standard error of prediction lower than 1.55%.
15