Semiconductor Diodes and Transistors as Electrical Thermometers

A. G. McNamara

Review of Scientific Instruments · 1962 · 70 citations · 1 references

Concepts

TL;DR

Electrical thermometers for laboratory, environmental, and rocket‑borne applications can be built with simple semiconductor circuits. The devices use the temperature‑dependent voltage of a forward‑conducting p‑n junction; a diode with a microammeter measures this voltage for lab use, while a single‑transistor, three‑resistor circuit amplifies the base‑emitter voltage for airborne telemetry, with adjustable resistors setting range and sensitivity. The forward‑junction voltage changes by about 2 mV/°C, remaining essentially linear from −40 °C to +100 °C.

Abstract

Electrical thermometers for laboratory experiments, environmental testing, and flight performance of rocket-borne equipment may be realized with simple semiconductor circuits. The devices described here utilize the variation with temperature of voltage across a forward-conducting p-n junction. This voltage variation of approximately 2 mv/°C is virtually linear over the entire test range of −40 to +100°C. A diode used in conjunction with a microammeter to measure the voltage across the diode is most convenient for laboratory use. For airborne use, where a signal varying from 0 to 5 v is required for telemetry, the transistor circuit is ideal. The latter circuit, which amplifies its own base-emitter junction voltage variations, requires only a single transistor and 3 resistors. Adjustment of two of the resistance values enables the user to select any temperature range to be covered with the desired sensitivity.

References

1