Analytical Chemistry · 2013 · 27 citations · 23 references
EngineeringBioengineeringBiomedical EngineeringBiomemsPrecision MedicinePrecision OncologyCirculating Tumor CellsFluid AnalysisMicrofluidic Device FabricationMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsCell Sorting DevicesCell Sorting In Microfluidic DevicesMicrofluidic AnalysisMicrofluidic Cell SortingMicrofluidic Chip DesignMicrofluidic PlatformsMicrofluidic SortingSorting
Ensemble-decision aliquot ranking (eDAR) is a sensitive and high-throughput method to analyze circulating tumor cells (CTCs) from peripheral blood. Here, we report the next generation of eDAR, where we designed and optimized a new hydrodynamic switching scheme for the active sorting step in eDAR, which provided fast cell sorting with an improved reproducibility and stability. The microfluidic chip was also simplified by incorporating a functional area for subsequent purification using microslits fabricated by standard lithography method. Using the reported second generation of eDAR, we were able to analyze 1 mL of whole-blood samples in 12.5 min, with a 95% recovery and a zero false positive rate (n = 15).
23
Isolation of rare circulating tumour cells in cancer patients by microchip technology
Sunitha Nagrath, Lecia V. Sequist, Shyamala Maheswaran et al. · Nature · 2007 · 3.5K citations · Full text
Medicine, Cancer Biology, Oncology +15