Electroanalysis · 2014 · 12 citations · 30 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectroanalytical ChemistryElectrochemical DetectionEnvironmental AnalysisEnvironmental Analytical ChemistryCarbon-based Working ElectrodesChromium Analytical DetectionChromium Speciation AnalysisMethod Detection LimitPreconcentrationSample PreconcentrationVoltammetric DetectionVoltammetric Trace Analysis
Abstract Pyrolyzed photoresist carbon electrodes modified with bismuth (Bi‐PPCEs) were prepared and used for the determination of trace levels of chromium. In order to lower the detection limit, the choice of the supporting electrolyte concentration and electroanalytical parameters were optimized. The adsorptive cathodic stripping voltammetric determination of Cr(VI) at Bi‐PPCE was selective even in the presence in Cr(III). A detection limit of 0.1 µg L −1 was obtained after 60 s preconcentration. After oxidation with UV light and oxygen, all chromium species were oxidized to Cr(VI) and the total inorganic Cr determination is possible. Finally, Bi‐PPCE was applied to trace Cr(VI) analysis in real samples such as waste water and tap water.
30
Bismuth-Coated Carbon Electrodes for Anodic Stripping Voltammetry
Joseph Wang, Jianmin Lu, Samo B. Hočevar et al. · Analytical Chemistry · 2000 · 998 citations
Melissa A. Nolan, Samuel P. Kounaves · Analytical Chemistry · 1999 · 208 citations
C-MEMS for the Manufacture of 3D Microbatteries
Chunlei Wang, Lili Taherabadi, Guangyao Jia et al. · Electrochemical and Solid-State Letters · 2004 · 201 citations