EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsBiomedical Laser ApplicationsOptical CoherenceOptical Coherence TomographyOptical ImagingOptical TomographyTissue OpticsAxial ResolutionBroadband Light SourcesCatheter-based Imaging DevicesChromatic Aberration CorrectionEndoscopic Optical ImagingLateral ResolutionOptical Coherence Tomography InstrumentationSpectral-domain Optical Coherence Tomography
We present a novel design for an endoscopic imaging catheter utilizing diffractive optics for ultrahigh-resolution optical coherence tomography (OCT) imaging at 800 nm. A diffractive microlens was developed to alleviate severe chromatic aberration when a broadband light source was employed at the 800 nm wavelength range. Combined with a home-built fiber rotary joint and a broadband Ti:sapphire laser, the imaging catheter achieved a lateral resolution of 6.2 μm and an axial resolution of 3.0 μm in air. The performance of the catheter was demonstrated by three-dimensional full-circumferential endoscopic OCT imaging of guinea pig esophagus in vivo.
11
Ultrahigh-resolution ophthalmic optical coherence tomography
Wolfgang Drexler, Uwe Morgner, Ravi K. Ghanta et al. · Nature Medicine · 2001 · 959 citations · Full text
Imaging needle for optical coherence tomography
Xingde Li, C. Chudoba, Tony H. Ko et al. · Optics Letters · 2000 · 212 citations