Applied Optics · 2000 · 26 citations · 18 references
EngineeringPhysicsOptical MaterialsOpticsVisual OpticsAdaptive OpticsCalibrationOptical AberrationsOptical InstrumentationOptical MetrologySensor CalibrationAdaptive Optics CalibrationAdaptive Optics InstrumentationDeformable MirrorsOptical Wavefront SensingStrehl RatioWavefront Error EstimationWavefront MetrologyWavefront Sensing
Calibration experiments with a bimorph mirror are presented. Phase-diversity wave-front sensing is used for measuring the control matrix, nulling wave-front errors in the optical setup, including the mirror, and measuring Strehl ratios and residual higher-order aberrations. The Strehl ratio of the calibrated system is measured to be 0.975, corresponding to 1/40 wave rms in the residual wave front. The conclusion is that a phase-diversity wave-front sensor is easier to install and use than interferometers and can replace them in optical setups for testing adaptive optics systems.
18