Multiple Module Pixellated CdTe Spectroscopic X-Ray Detector

Matthew D. Wilson, Steven Bell, Robert J. Cernik, Christiana Christodoulou, Christopher K. Egan, Daniel O’Flynn, Simon D. M. Jacques, S. Pani, James Scuffham, P. Seller,

IEEE Transactions on Nuclear Science · 2013 · 30 citations · 16 references

Concepts

Abstract

A pixellated CdTe detector system comprising 2×2 detector modules has been developed for high energy spectroscopic X-ray imaging applications and has an active area of 16 cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The detector modules are made from the HEXITEC 80×80 ASIC and 1 mm thick CdTe with Al-Schottky contacts. The CdTe has 250 μm pitch pixels with an outer guard ring on the same pitch. The single HEXITEC 80×80 detectors have an average energy resolution (FWHM) of 800 eV at 59.9 keV. Limitations in the multiple module DAQ system mean that the energy resolution of the pixels in the 2×2 detector array is 2.0 keV at 59.9 keV. The spacing between the tiled detector modules is 150 μm which results in an inactive area equivalent to 3 pixels, including the guard ring on the edge of the detectors. The modular detector configuration demonstrates the potential to create large area detector arrays in the future.