IEEE Transactions on Nuclear Science · 2013 · 30 citations · 16 references
EngineeringElectrical EngineeringElectronic MaterialsRadiation DetectionSpectroscopyDetector TechnologyEnergy ResolutionRadiation DetectorsSemiconductor DetectorsX-ray DetectionX-ray DetectorsDetector ArraysDetector DesignDetector Energy ResolutionHigh-resolution X-ray DetectorsPixelated Detector ArraysRadiation Energy ResolutionSchottky ContactsX-ray Spectroscopy Detectors
A pixellated CdTe detector system comprising 2×2 detector modules has been developed for high energy spectroscopic X-ray imaging applications and has an active area of 16 cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The detector modules are made from the HEXITEC 80×80 ASIC and 1 mm thick CdTe with Al-Schottky contacts. The CdTe has 250 μm pitch pixels with an outer guard ring on the same pitch. The single HEXITEC 80×80 detectors have an average energy resolution (FWHM) of 800 eV at 59.9 keV. Limitations in the multiple module DAQ system mean that the energy resolution of the pixels in the 2×2 detector array is 2.0 keV at 59.9 keV. The spacing between the tiled detector modules is 150 μm which results in an inactive area equivalent to 3 pixels, including the guard ring on the edge of the detectors. The modular detector configuration demonstrates the potential to create large area detector arrays in the future.