Natural SciencesMaterials EngineeringMaterials ScienceSurface ChemistrySurface ScienceThin FilmsSurface HydrophobicitySurface TensionThin Film Mechanical PropertiesForce-distance CurvesLangmuir MonolayersLinear Variable Differential TransformersMonolayer FilmsMonolayersSurface PressureThin Film Recording MediaThin Film Surface Characterization
Trurnit and Lauer [Rev. Sci. Instr. 30, 975 (1959)] have recently described an automatic recording film balance system suitable for automatic recording of many properties of insoluble surface films. The authors have been working with an instrument similar in principle, somewhat less versatile, but very simply constructed by modifying a Cenco hydrophil balance. It is well suited for automatic recording of force-area curves for monolayers spread on aqueous substrates. Major components include a linear differential transformer, lead screw barrier advance system, and an x-y recorder. The practicality of the appratus was demonstrated on a stearic acid monolayer spread from benzene on a distilled water substrate. Sensitivity of 0.01 d/cm in spreading pressure and 0.1 A2 in molecular co-area was readily obtained, together with a surface pressure range of 0–60 d/cm.