Review of Scientific Instruments · 1959 · 25 citations · 2 references
EngineeringElectrical EngineeringEquipmentInstrumentationSensorsPressure SensorsProcess InstrumentationRecording TechnologyAnalytical BalancesLinear Variable Differential TransformersViscosityVolta Potential
An automatic recording film balance system is described. It consists of a film trough and two pendulum type balances with pressure ranges of 0–400 (800) millidynes/cm and 0–40 (80) dynes/cm, respectively. Each balance is connected to a float when in operation (Langmuir type balance) and the sensing element in both cases is a linear variable differential transformer, the iron core of which is connected to the pendulum. The transformer signal after amplification drives the pen motor of a strip chart recorder. Special features like a constant pressure device together with auxiliary instruments such as a second X-Y recorder, an automatic dipping device for film transfer, a volta potential meter and a slit viscosimeter permit the automatic recording of the following functions: (dFdA)T, (dAdT)F, (dFdT)A, (dFdt)T, A, (dAdt)T, F, (AwAs)T, F, (dVdF)T, and (dηdF)T.F=film pressure, A=film area, T=temperature, t=time, Aw/As=deposition ratio, η=film viscosity, and V=volta potential. Test and performance data are presented.
2
A theory and method for the spreading of protein monolayers
H. J. Trurnit · Journal of Colloid Science · 1960 · 217 citations