Review of Scientific Instruments · 1952 · 10 citations · 0 references
EngineeringChemical EngineeringChemistryAnalytical ChemistryInstrumentationThin FilmsPressure LoggingPressure MeasurementPressure MetrologyCompressionExpansionSurface Pressure - Area Isotherms
A null-deflection film balance which retains the sensitivity and range of the manually operated Langmuir-Adam balance but which is capable of following rate processes continuously is described. It provides automatic plotting of (1) film pressure as a function of molecular area during continuous compression or expansion, (2) film pressure as a function of time at constant film area, and (3) film area as a function of time at constant film pressure. The design utilizes commercially available components. Recording is done with a pen-and-ink recorder of standard type, and the curves are visible in their entirety during an experiment. The coordinates are adjustable to make the chart direct reading in F(A), F(t), or A(t) to any convenient scale or degree of scale expansion. Continuous use for 18 months has shown the instrument to be consistently reliable and convenient.