Optics Express · 2013 · 100 citations · 26 references
EngineeringRobotics EngineeringInstrumentationMetrologyRobot KinematicsAngle MeasurementAngular MeasurementCalibrationGeometric MeasurementMeasurement ErrorSensor CalibrationAngle Error AnalysisAngle Measurement Error SourcesAngular Measurement SystemsAngular MetrologyGeometric Error CompensationGeometric Error MeasurementLaser Metrology SystemsStraightness Measurement
A simple method for simultaneously measuring the 6DOF geometric motion errors of the linear guide was proposed. The mechanisms for measuring straightness and angular errors and for enhancing their resolution are described in detail. A common-path method for measuring the laser beam drift was proposed and it was used to compensate the errors produced by the laser beam drift in the 6DOF geometric error measurements. A compact 6DOF system was built. Calibration experiments with certain standard measurement meters showed that our system has a standard deviation of 0.5 µm in a range of ± 100 µm for the straightness measurements, and standard deviations of 0.5", 0.5", and 1.0" in the range of ± 100" for pitch, yaw, and roll measurements, respectively.
26
Error Compensation of Coordinate Measuring Machines
G. Zhang, R. Veale, T. Charlton et al. · CIRP Annals · 1985 · 238 citations