Experimental determination of the transmission factor for the Omicron EA125 electron analyzer
33
Citations
6
References
2000
Year
EngineeringPositron Annihilation SpectroscopyMeasurementIdeal AnalyzerOmicron Ea125Optical CharacterizationTransmission FactorElectron OpticElectron SpectroscopyOptical DiagnosticsInstrumentationElectrical EngineeringExperimental DeterminationAccelerator Mass SpectrometryPhysicsConstant Analyzer EnergyAtomic PhysicsUniversal LensNatural SciencesSpectroscopyInstrument ScienceApplied Physics
In this article a study of the transmission factor of the Omicron EA125 analyzer equipped with the universal lens is presented. The procedure is based on a model by Cross and Castle [J. Electron Spectrosc. Relat. Phenom. 22, 53 (1981)] and is applicable to every spectrometer which can be operated in the constant analyzer energy (CAE) and in the constant retarding ratio measuring mode. The advantage of the method is its independence on the sample and on the inelastic mean free path of the electrons. We find that the transmission factor for the CAE mode is proportional to Ekin−1 for most measuring setups. This dependence is predicted by theory for an ideal analyzer. Deviations from this behavior are observed if the retarding ratio for a given kinetic energy is too small. The limit value of the retarding ratio for ideal behavior, i.e., an Ekin−1 transmission factor, depends on the analyzer entrance slit aperture which has been selected.