Nanotechnology · 2014 · 26 citations · 45 references
EngineeringMechanical EngineeringActuatorsBiomechanicsVibrationAtomic Force MicroscopyContact AnalysisContact DynamicsContact MechanicsScanning Probe MicroscopyCantilever MechanicsDampingFluid LoadingLiquid Surface ResonanceStiffnessStiffness MeasurementSurface StiffnessViscous Damping
Contact resonance AFM (CR-AFM) is a scanning probe microscopy technique that utilizes the contact resonances of the AFM cantilever for concurrent imaging of topography and surface stiffness. The technique has not been used in liquid until recently due to analytical and experimental difficulties, associated with viscous damping of cantilever vibrations and fluid loading effects. To address these difficulties, (i) an analytical approach for contact resonances in liquid is developed, and (ii) direct excitation of the contact resonances is demonstrated by actuating the cantilever directly in a magnetic field. By implementing the analytical approach and the direct actuation through magnetic particles, quantitative stiffness imaging on surfaces with a wide range of stiffness can be achieved in liquid with soft cantilevers and low contact forces.
45
Calibration of atomic-force microscope tips
Jeffrey L. Hutter, John Bechhoefer · Review of Scientific Instruments · 1993 · 4.2K citations
Short cantilevers for atomic force microscopy
D. A. Walters, J. P. Cleveland, Neil H. Thomson et al. · Review of Scientific Instruments · 1996 · 490 citations