Micro‐electrospray with stainless steel emitters

Wenqing Shui, Yanling Yu, Xuejiao Xu, Zhenyu Huang, Guobing Xu, Pengyuan Yang

Rapid Communications in Mass Spectrometry · 2003 · 27 citations · 23 references

Concepts

TL;DR

Wettability and hydrophobicity of stainless steel tips are thought to determine micro‑electrospray performance. The study investigated the physical processes governing micro‑electrospray performance using a blunt‑tip stainless steel emitter. The authors reduced the emitter orifice to 50–150 nL/min, enabling attomole‑level peptide detection, and incorporated the stainless steel emitter into a PMMA microchip for efficient analysis of a myoglobin digest. Optimized ESI conditions allowed sheathless micro‑electrospray from a blunt stainless steel tip, with a shrinking Taylor cone enabling a 150 nL/min flow rate and attomole‑level peptide detection, and the emitter performed well in a PMMA microchip for myoglobin digest analysis.

Abstract

The physical processes underlying micro-electrospray (micro-ES) performance were investigated using a stainless steel (SS) emitter with a blunt tip. Sheathless micro-ES could be generated at a blunt SS tip without any tapering or sanding if ESI conditions were optimized. The Taylor cone was found to shrink around the inner diameter of the SS tubing, which permitted a low flow rate of 150 nL/min for sheathless microspray on the blunt tip (100 microm i.d. x 400 microm o.d.). It is believed that the wettability and/or hydrophobicity of SS tips are responsible for their micro-ES performance. The outlet orifice was further nipped to reduce the size of the spray cone and limit the flow rate to 50-150 nL/min, resulting in peptide detection down to attomole quantities consumed per spectrum. The SS emitter was also integrated into a polymethylmethacrylate microchip and demonstrated satisfactory performance in the analysis and identification of a myoglobin digest.

References

23