Simultaneous Measurement of Hydrocarbons and Sulfur Compounds using Flame Ionization and Sulfur Chemiluminescence Detection for Sulfur Simulated Distillation

Randy L. Shearer, Linda M. Meyer

Journal of High Resolution Chromatography · 1999 · 14 citations · 12 references

Concepts

Abstract

A gas chromatographic system for the simultaneous acquisition of hydrocarbon and sulfur chromatograms was developed. Detection of sulfur compounds is achieved using a sulfur chemiluminescence detector (SCD) mounted in series with a flame ionization detector (FID). A constant fraction of the effluent of the FID is transferred to the SCD by means of a fixed restrictor. Unlike previous versions of this approach, the FID is not used to generate the chemiluminescent sulfur species. Rather, the FID is operated under optimum conditions for hydrocarbon analysis and a furnace is used to generate the chemiluminescent sulfur species. The system permits dual acquisition of the hydrocarbon and sulfur signals in a single analysis with a single column, since the detectors are operated in a serial fashion. The application of sulfur simulated distillation using this approach was examined, since this requires simultaneous universal and sulfur selective detection. Precision of absolute response of both the FID and SCD was typically less than 2% RSD for a standard reference material.

References

12