Journal of High Resolution Chromatography · 1999 · 14 citations · 12 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisDetector SystemsGas ChromatographyInstrumental AnalysisSulfur ChemistrySulfur Compound AnalysisSulfur CompoundsFlame Ionization DetectionGas Chromatography DetectorsGas Detection InstrumentsHydrocarbon DetectionSulfur Compound DetectionSulfur Compound QuantificationSulfur DeterminationVolatile Sulfur Compounds
A gas chromatographic system for the simultaneous acquisition of hydrocarbon and sulfur chromatograms was developed. Detection of sulfur compounds is achieved using a sulfur chemiluminescence detector (SCD) mounted in series with a flame ionization detector (FID). A constant fraction of the effluent of the FID is transferred to the SCD by means of a fixed restrictor. Unlike previous versions of this approach, the FID is not used to generate the chemiluminescent sulfur species. Rather, the FID is operated under optimum conditions for hydrocarbon analysis and a furnace is used to generate the chemiluminescent sulfur species. The system permits dual acquisition of the hydrocarbon and sulfur signals in a single analysis with a single column, since the detectors are operated in a serial fashion. The application of sulfur simulated distillation using this approach was examined, since this requires simultaneous universal and sulfur selective detection. Precision of absolute response of both the FID and SCD was typically less than 2% RSD for a standard reference material.
12
Pulsed Flame Photometer Detector for Gas Chromatography
Aviv Amirav, Hongwu Jing · Analytical Chemistry · 1995 · 106 citations