Review of Scientific Instruments · 1958 · 49 citations · 5 references
Natural SciencesNuclear EngineeringRadiation DetectionRadiation MeasurementSpectroscopyRadiation SpectroscopyX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray OpticsX-ray SpectroscopyInput Data AnalysisPhotopeak AnalysisResponse Matrix AnalysisResponse Matrix ModelingSodium Iodide DetectorsX-ray Imaging PhysicsX-ray Spectral AnalysisX-ray Spectroscopy Detectors
The response of a four-inch long by five-inch diameter sodium-iodide crystal to axially incident 0.01- to 8-Mev x-rays has been expressed as a 28 by 28 matrix. Measured pulse-height distributions were combined with Monte Carlo calculated distributions for the input data. The matrix was inverted on an automatic computer for ease of application to spectra. The mesh is uniform in the square root of energy, each interval including an entire photopeak. Such a coarse mesh makes the response matrix suitable for the unscrambling of continuous but not line spectra.
5