Chemical Engineering Communications · 1986 · 67 citations · 9 references
Life SciencesBiochemistryMicrobiologyCellular MetabolismMetabolismMicrobial MetabolismMetabolic Pathway AnalysisMetabolic Pathways In FermentationYeast BiologyConcentrationFermentation PathwaysNadh In FermentationNadh MeasurementNadh QuantificationPrimary Metabolism During FermentationYeast Fermentation PhysiologyYeast Metabolism
Abstract Metabolic activity contributions to culture fluorescence were studied by observing the FluroMeasure Detector response to pulse additions of various carbon sources to yeast fermentations. The shape of the resulting pulses in the fluorometer signal is characteristic of the C-source and amount added. The pulse height is a function of the cell concentration, and the pulse area is a function of the amount of C-source added. Specific features of the signal response are related to specific steps in the primary metabolic pathways in yeast. Measurements of culture fluorescence derived from intracellular NADH, therefore, provide on-line information on the primary metabolism within yeast. Since the primary metabolic pathways in yeast are similar to those used in many procaryotic and higher eucaryotic cells, the measurement of intracellular NADH is likely to provide similar information to monitor and eventually control fermentation processes involving a wide variety of microbial or cell cultures.
9