Improved estimation of fructosamine, as a measure of glycated serum protein, with the Technicon RA-1000 analyzer.

Fernando San-Gil, G M Schier, Robert G. Moses, I.E.T. Gan

Clinical Chemistry · 1985 · 35 citations · 0 references

Concepts

Abstract

We present an improved fructosamine assay for use with the Technicon RA-1000 analyzer. Features of this assay include: specimen throughput of 100 per hour, within-batch CV of 1.5%, between-batch CV of 2.4%, standard curve linear up to 24.9 mmol of fructosamine per liter, and good correlation (r = 0.82) with hemoglobin A1.