Offline impedance measurements for detection and mitigation of dangerous implant interactions: An RF safety prescreen

Christopher W. Ellenor, Pascal Stang, Maryam Etezadi‐Amoli, John M. Pauly, Greig Scott

Magnetic Resonance in Medicine · 2014 · 12 citations · 22 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: The concept of a "radiofrequency safety prescreen" is investigated, wherein dangerous interactions between radiofrequency fields used in MRI, and conductive implants in patients are detected through impedance changes in the radiofrequency coil. THEORY: The behavior of coupled oscillators is reviewed, and the resulting, observable impedance changes are discussed. METHODS: A birdcage coil is loaded with a static head phantom and a wire phantom with a wire close to its resonant length, the shape, position, and orientation of which can be changed. Interactions are probed with a current sensor and network analyzer. RESULTS: Impedance spectra show dramatic, unmistakable splitting in cases of strong coupling, and strong correlation is observed between induced current and scattering parameters. CONCLUSIONS: The feasibility of a new, low-power prescreening technique has been demonstrated in a simple phantom experiment, which can unambiguously detect resonant interactions between an implanted wire and an imaging coil. A new technique has also been presented which can detect parallel transmit null modes for the wire.

References

22