Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating
232
Citations
8
References
2004
Year
The findings revealed marked differences across two MR systems in the level of radiofrequency (RF)-induced temperature changes per unit of whole body SAR for a conductive implant. Thus, these data suggest that using SAR to guide MR safety recommendations for neurostimulation systems or other similar implants across different MR systems is unreliable and, therefore, potentially dangerous. Better, more universal, measures are required in order to ensure patient safety.