Journal of Magnetic Resonance Imaging · 2000 · 297 citations · 6 references
Natural SciencesBiomedical EngineeringRadiologyDiagnostic ImagingHeat TransferMedical PhysicsInterventional Radiology PhysicsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsRadio-frequency HeatingRadiofrequency ExcitationEnvironmental FactorsInterventional Magnetic Resonance ImagingMagnetic Resonance-guided Intervention PhysicsSafety In Magnetic Resonance ImagingTemperature Rise AnalysisTemperature Rise TestingThermal Tissue Effects
We examined the unwanted radiofrequency (RF) heating of an endovascular guidewire frequently used in interventional magnetic resonance imaging (MRI). A Terumo guidewire was partly immersed in an oblong saline bath to simulate an endovascular intervention. The temperature rise of the guidewire tip during an FFE sequence [average specific absorption rate (SAR) = 3.9 W/kg] was measured with a Luxtron fluoroscopic fiber. Starting from 26 degrees C, the guidewire tip reached temperatures up to 74 degrees C after 30 seconds of scanning. Touching the guidewire may cause sudden heating at the point of contact, which in one instance caused a skin burn. The excessive heating of a linear conductor like the guidewire can only be explained by resonating RF waves. The capricious dependencies of this resonance phenomenon on environmental factors have severe consequences for predictability and safety guidelines.
6
Recording of EEG during fMRI experiments: Patient safety
Louis Lemieux, Philip J. Allen, Florence Franconi et al. · Magnetic Resonance in Medicine · 1997 · 312 citations
Magnetic Resonance Bioeffects, Safety and Patient Management
Journal of Neuroimaging · 1995 · 155 citations
Active MR visualization of a vascular guidewire in vivo
Mark E. Ladd, G Zimmermann, Harald H. Quick et al. · Journal of Magnetic Resonance Imaging · 1998 · 98 citations
Chris J.G. Bakker, Henk F.M. Smits, Clemens Bos et al. · Journal of Magnetic Resonance Imaging · 1998 · 63 citations · Full text