The Analyst · 2015 · 37 citations · 37 references
EngineeringBioengineeringBiomedical EngineeringAnalytical ChemistryMicroelectromechanical SystemsMicrofabricationMagnetic Resonance SpectroscopyMicrofluidic BiosensorsMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsMicroscale BiosensorsMicrofluidic Device DesignMicrofluidic Diagnostic DevicesMicrofluidic Diagnostic PlatformsMicrofluidic SensingMicroscale SensingSample Preparation MicrodevicesTransceiver ArchitecturesTransceiver Design
Herein, we describe a micro-nuclear magnetic resonance (μNMR) relaxometer miniaturized to palm-size and electronically automated for multi-step and multi-sample chemical/biological diagnosis. The co-integration of microfluidic and microelectronic technologies enables an association between the droplet managements and μNMR assays inside a portable sub-Tesla magnet (1.2 kg, 0.46 Tesla). Targets in unprocessed biological samples, captured by specific probe-decorated magnetic nanoparticles (NPs), can be sequentially quantified by their spin-spin relaxation time (T2) via multiplexed μNMR screening. Distinct droplet samples are operated by a digital microfluidic device that electronically manages the electrowetting-on-dielectric effects over an electrode array. Each electrode (3.5 × 3.5 mm(2)) is scanned with capacitive sensing to locate the distinct droplet samples in real time. A cross-domain-optimized butterfly-coil-input semiconductor transceiver transduces between magnetic and electrical signals to/from a sub-10 μL droplet sample for high-sensitivity μNMR screening. A temperature logger senses the ambient temperature (0 to 40 °C) and a backend processor calibrates the working frequency for the transmitter to precisely excite the protons. In our experiments, the μNMR relaxometer quantifies avidin using biotinylated Iron NPs (Φ: 30 nm, [Fe]: 0.5 mM) with a sensitivity of 0.2 μM. Auto-handling and identification of two targets (avidin and water) are demonstrated and completed within 2.2 min. This μNMR relaxometer holds promise for combinatorial chemical/biological diagnostic protocols using closed-loop electronic automation.
37
Alexander D. Douglas, Andrew R. Williams, Joseph J. Illingworth et al. · Nature Communications · 2013 · 926 citations · Full text
Magnetic relaxation switches capable of sensing molecular interactions
Juan Manuel Sánchez Pérez, Lee Josephson, T. O'Loughlin et al. · Nature Biotechnology · 2002 · 892 citations · Full text
Chip–NMR biosensor for detection and molecular analysis of cells
Hakho Lee, Eric Sun, Donhee Ham et al. · Nature Medicine · 2008 · 625 citations · Full text
Development of a digital microfluidic platform for point of care testing
Ramakrishna Sista, Zhishan Hua, Prasanna K. Thwar et al. · Lab on a Chip · 2008 · 542 citations · Full text