IEEE Symposium on Ultrasonics · 2002 · 29 citations · 6 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyUltrasoundDiagnostic Imaging SystemsImage ReconstructionImagingMedical ImagingMedical Imaging PhysicsUltrasound ImagingB-mode UltrasonographyB-scan ImagingImage Contrast EvaluationImage Contrast MechanismsScatteringSparse Array DesignTwo-dimensional ArraysVolume Scattering
The potential of using sparse 2-D arrays for medical imaging is assessed by simulating B-scan images of spherical nonscattering cysts embedded in a random scattering volume. The imaging potential of an eighth-order sparse array having close to 1000 elements is demonstrated by quantitatively comparing image contrast between a nonscattering cyst and the surrounding random scattering medium. A 32nd-order sparse array with 245 elements is found to perform at a reduced level, producing unacceptable artifactual echoes within images of cysts. It is concluded that 2-D arrays have the potential to produce medical images superior to linear phased arrays, as well as the ability to produce high-quality images of low-contrast objects at arbitrary steering angles.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
6
Computer Simulations of Speckle in B-Scan Images
David R. Foster, Marcel Arditi, F. Stuart Foster et al. · Ultrasonic Imaging · 1983 · 74 citations