Journal of Mass Spectrometry · 2012 · 210 citations · 23 references
Life SciencesBiochemistryAnalytical ChemistryLipid BiochemistryMass Spectrometric AnalysisLipid AnalysisLipidomicsMass AnalyzersMass SpectrometryMass Spectrometry Data AnalysisIonizationLipid Composition AnalysisLipid ProfilingLipid QuantificationMass Spectrometry-based LipidomicsOrbitrap Mass AnalyzersOrbitrap Mass SpectrometryTandem Mass Spectrometry
Top-down shotgun lipidomics relies on direct infusion of total lipid extracts into a high-resolution tandem mass spectrometer and implies that individual lipids are recognized by their accurately determined m/z. Lipid ionization efficiency and detection specificity strongly depend on the acquisition polarity, and therefore it is beneficial to analyze lipid mixtures in both positive and negative modes. Hybrid LTQ Orbitrap mass spectrometers are widely applied in top-down lipidomics; however, rapid polarity switching was previously unfeasible because of the severe and immediate degradation of mass accuracy. Here, we report on a method to rapidly acquire high-resolution spectra in both polarity modes with sub-ppm mass accuracy and demonstrate that it not only simplifies and accelerates shotgun lipidomics analyses but also improves the lipidome coverage because more lipid classes and more individual species within each class are recognized. In this way, shotgun analysis of total lipid extracts of human blood plasma enabled to quantify 222 species from 15 major lipid classes within 7 min acquisition cycle.
23