Journal of Refractive Surgery · 1991 · 29 citations · 14 references
Health ProfessionsOphthalmologyVision ScienceClinical Vision ScienceOcular AnatomyOcular ImagingCorneal AnatomyCorneal ImagingCorneal OpticsCorneal TopographyOcular BiometryOcular Surface ImagingAssessment Error AnalysisComputational Reconstruction AlgorithmsCorneal Astigmatism AnalysisCorneal Topography AssessmentCorneal Topography InterpretationImage Reconstruction AlgorithmsMisalignment EffectsReconstruction Algorithms
This article presents a general method of error analysis for reconstruction algorithms employed by corneal topography systems and applies this method to assess the accuracy of the Wang reconstruction algorithms. Using mathematical surface reconstructions, the method detects artifacts introduced in topographic analysis by astigmatism and misalignment. For the Wang algorithms, error introduced by astigmatism was less than 1% of the true value for surfaces with more than 10.00 diopters of astigmatism. A 1-mm translational misalignment of the analyzed surface introduced error of no more than 1.3%. Errors caused by rotational misalignment were less than 3% for misalignments of 5 degrees or less. It is concluded from the error analysis method that the Wang algorithms are sufficiently accurate for use in the analysis of corneal topography. Similar analyses need to be accomplished for the validation of commercially available corneal topography systems.
14
J. James Rowsey · Archives of Ophthalmology · 1981 · 130 citations
Medicine, Ophthalmology, Surgery +17
Dennis J. Gormley, Martin Gersten, Richard S. Koplin et al. · Cornea · 1988 · 100 citations
Medicine, Ophthalmology, Surgery +17
Method for Calculation of Corneal Profile and Power Distribution
J.D. Doss, R. L. Hutson, J. James Rowsey et al. · Archives of Ophthalmology · 1981 · 93 citations