PubMed · 1989 · 44 citations · 0 references
MedicineHematologySurgeryBiocompatibilityHemostasisVascular SurgeryExtracorporeal CirculationThrombosisThrombosis PreventionCircuit CoatingExtracorporeal Membrane Oxygenation AnticoagulationHeparin AnticoagulationSystemic Anticoagulation
Adult sized extracorporeal membrane oxygenation circuits were coated with Duraflo II chemical heparin bonding by Baxter-Bentley Laboratories. Five sheep were maintained on venovenous extracorporeal circulation for four days with no systemic anticoagulation. There was no bleeding, no major thrombosis in the circuits, and no significant emboli after 4 days of extracorporeal circulation without anti-coagulation.