PharmacyBiochemistryAnalytical ToxicologyBiopharmaceuticsMetabolic StabilityBioanalytical Method ValidationClinical AnalysisRenal Function TestsAssayColorimetric TestsCreatinine Analysis In Body FluidsCreatinine Assay InterferenceLinear RangePrecisionSerum Creatinine AssessmentTest Performance Characteristics
We evaluated the performance characteristics of the Beckman Creatinine Analyzer 2, a kinetic colorimetric analyzer which utilizes a modified Jaffé reagent. The instrument gave linear measurements to at least 300 mg/L and its within-run precision on a 50 mg/l creatinine standard was 1.1%. Its between-day precision on the 50 mg/L creatinine standard was less than 5.0%. Its results correlated well with those from the manual Jaffé. CentrifiChem System 400, TechniconTM SMATM 6/60 System, and TraceTM systems I and III. Interferences from 19 possible serum constituents and 32 drugs were almost nonexistent. The extended linear range of the Creatinine Analyzer 2 allowed the direct assay of samples from patients undergoing periodic hemodialysis.