IEEE Transactions on Biomedical Circuits and Systems · 2024 · 12 citations · 49 references
EngineeringBioengineeringBiomedical EngineeringBioinstrumentationMicroelectromechanical SystemsNeural EngineeringBiomedical InstrumentationImplantable Biomedical DevicesImplantable MicrodevicesImplantable MicrosystemsImplantable Neural DevicesNeural InterfacesBendingBending StiffnessImplantable Neural ElectrodesImplantable Neural InterfacesNeural Electrode ArraysNeural Interface DesignNeural RecordingNeural Recording Systems
This article presents the system architecture for an implant concept called NeuroBus. Tiny distributed direct digitizing neural recorder ASICs on an ultra-flexible polyimide substrate are connected in a bus-like structure, allowing short connections between electrode and recording front-end with low wiring effort and high customizability. The small size (344 μm × 294 μm) of the ASICs and the ultraflexible substrate allow a low bending stiffness, enabling the implant to adapt to the curvature of the brain and achieving high structural biocompatibility. We introduce the architecture, the integrated building blocks, and the post-CMOS processes required to realize a NeuroBus, and we characterize the prototyped direct digitizing neural recorder front-end as well as polyimide-based ECoG brain interface. A rodent animal model is further used to validate the joint capability of the recording front-end and thin-film electrode array.
49
NeuroGrid: recording action potentials from the surface of the brain
Dion Khodagholy, Jennifer N. Gelinas, Thomas Thesen et al. · Nature Neuroscience · 2014 · 954 citations · Full text
A. Bolu Ajiboye, Francis R. Willett, Daniel R. Young et al. · The Lancet · 2017 · 916 citations · Full text
Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural Dust
Dongjin Seo, Ryan Neely, Konlin Shen et al. · Neuron · 2016 · 559 citations · Full text