High-resolution and high-sensitivity X-ray ptychographic coherent diffraction imaging using the CITIUS detector

Yukio Takahashi, Masaki Abe, Hideshi Uematsu, Shuntaro Takazawa, Yuhei Sasaki, Nozomu Ishiguro, Kyosuke Ozaki, Yoshiaki Honjo, H. Nishino, Kazuo Kobayashi,

Journal of Synchrotron Radiation · 2023 · 18 citations · 33 references

DOIFull text

Open access

Concepts

Abstract

Ptychographic coherent diffraction imaging (PCDI) is a synchrotron X-ray microscopy technique that provides high spatial resolution and a wide field of view. To improve the performance of PCDI, the performance of the synchrotron radiation source and imaging detector should be improved. In this study, ptychographic diffraction pattern measurements using the CITIUS high-speed X-ray image detector and the corresponding image reconstruction are reported. X-rays with an energy of 6.5 keV were focused by total reflection focusing mirrors, and a flux of ∼2.6 × 10 10 photons s −1 was obtained at the sample plane. Diffraction intensity data were collected at up to ∼250 Mcounts s −1 pixel −1 without saturation of the detector. Measurements of tantalum test charts and silica particles and the reconstruction of phase images were performed. A resolution of ∼10 nm and a phase sensitivity of ∼0.01 rad were obtained. The CITIUS detector can be applied to the PCDI observation of various samples using low-emittance synchrotron radiation sources and to the stability evaluation of light sources.

References