IPMC Based Biosensor for the Detection of Biceps Brachii Muscle Movements

Kiwon Park, Bonghwan Lee, Hyung-Man Kim, Kap-Seung Choi, Gun-Yong Hwang, Gi–Sig Byun, Hyung-Ki Lee

International Journal of Electrochemical Science · 2013 · 15 citations · 14 references

DOIFull text

Open access

Concepts

Abstract

Ionic polymer-metal composite (IPMC) consists of a polyelectrolyte film both-surfaces plated by metallic electrodes. When the IPMC bends, a voltage is generated between the two electrodes across the membrane. Although the IPMC is a promising soft sensor applicable to biomedical application, it still requires a further signal processing technique to effectively measure the deflections. The present study investigated an IPMC sensor model based on a RC circuit that can be easily applicable to the detection of muscle movements. The RC circuit model, describing a physical composition of the IPMC, contains a current source that relates the amount of charge differences accumulated on each electrode to the applied bending angles. The effectiveness of the model was tested by comparing the reproduced bending angles from the inverse model with the real bending angles. Finally, the sensor system was applied to the detection of biceps brachii muscle movements that causes elbow flexion, and the results was compared with the conventional electromyography (EMG) method to identify the potential of the IPMC as a biomedical sensor.

References

14