Continuous Measurement of Lactate Concentration in Human Subjects through Direct Electron Transfer from Enzymes to Microneedle Electrodes

David M E Freeman, Damien Ming, Richard Wilson, Peter L. Herzog, C. Schulz, Alfons K. G. Felice, Yu‐Chih Chen, Danny O’Hare, Alison Holmes, Anthony E. G. Cass

ACS Sensors · 2023 · 55 citations · 29 references

DOIFull text

Open access

Concepts

Abstract

Microneedle lactate sensors may be used to continuously measure lactate concentration in the interstitial fluid in a minimally invasive and pain-free manner. First- and second-generation enzymatic sensors produce a redox-active product that is electrochemically sensed at the electrode surface. Direct electron transfer enzymes produce electrons directly as the product of enzymatic action; in this study, a direct electron transfer enzyme specific to lactate has been immobilized onto a microneedle surface to create lactate-sensing devices that function at low applied voltages (0.2 V). These devices have been validated in a small study of human volunteers; lactate concentrations were raised and lowered through physical exercise and subsequent rest. Lactazyme microneedle devices show good agreement with concurrently obtained and analyzed serum lactate levels.

References

29