Publication | Open Access
Treatment plan prescreening for patient-specific quality assurance measurements using independent Monte Carlo dose calculations
19
Citations
35
References
2022
Year
Independent dose calculation with the MC-based program ArcherQA has potential as a prescreen treatment for measurement-based patient-specific QA. AUC values (>0.9) showed excellent classification accuracy for monitoring failing plans with independent MC calculations.
| Year | Citations | |
|---|---|---|