IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119

Gary A. Ezzell, Jay Burmeister, Nesrin Dogan, T LoSasso, James Mechalakos, D Mihailidis, A Molineu, Jatinder Palta, Chester R. Ramsey, Bill J. Salter,

Medical Physics · 2009 · 1K citations · 18 references

DOIFull text

Open access

Concepts

TL;DR

IMRT commissioning is supported by credentialing tests from the Radiological Physics Center, and the developed protocol can serve as a benchmark for other facilities. Nine facilities evaluated IMRT planning and delivery using rectangular‑phantom test cases that incorporated ion‑chamber, film, and planar dosimetry with 3 %/3 mm gamma criteria, and the resulting mean and standard‑deviation data were used to establish confidence limits. Task Group 119 established quantitative baseline confidence limits for IMRT commissioning, and locally derived limits that exceed these values suggest a need for improved commissioning.

Abstract

AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept . Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.

References

18