Concepedia

Publication | Open Access

Application of a new volumetric microsampling device for quantitative bioanalysis of immunosuppression

12

Citations

33

References

2022

Year

Abstract

<b>Background:</b> Volumetric absorptive microsampling may reduce the blood collection burden associated with therapeutic drug monitoring of immunosuppression to prevent organ transplant rejection. This work describes the development of a laboratory and analytical technique for quantifying tacrolimus and mycophenolic acid (MPA) from the Tasso-M20™ in human whole blood using bead-based impact-assisted extraction. <b>Results:</b> The sampled blood volume was accurate with estimated volumes within <2% of the expected 20 μl. Recovery using impact-assisted extraction was 73-87% for MPA and 100% for tacrolimus and was hematocrit-independent for both analytes. The LC-MS/MS assay is precise and accurate within the acceptance criteria of 15%. <b>Conclusion:</b> The sampling and extraction procedures allowed for accurate quantification of tacrolimus and MPA. Exploration of abuse scenarios identified important education points for patients conducting home-based sample collections in the future.

References

YearCitations

Page 1