Clinical Chemistry · 1982 · 71 citations · 0 references
MedicinePharmaceutical AnalysisPharmacologyClinical PharmacologyNeuropharmacologyPsychopharmacologyAntidepressant PharmacologyAntidepressant PharmacotherapySerotonergic PharmacologyBlood Drug AnalysisConcentrationElectron Capture DetectionLinear RangePlasma Level AssaysPlasma SamplingPrecisionQuantitative Gas Chromatography
Abstract This gas-chromatographic method for assay of fluoxetine and norfluoxetine in human plasma involves extraction of the drugs and use of a 63Ni electron-capture detector. The linear range of detection is 25 to 800 micrograms/L for each drug. Overall precision (CV) in the concentration range of 10 to 100 micrograms/L for both drugs was approximately 10%. Accuracy (relative error) in the same concentration range was approximately +10%. None of the commonly prescribed antidepressants or tranquilizers that we tested interfere with the assay.