Physica Scripta · 2022 · 17 citations · 20 references
Natural SciencesPhysicsNuclear InstrumentationNuclear PhysicsSpectroscopyCalibrationDetector TechnologyPrecision SpectroscopyRadiation DetectorsX-ray DetectorsX-ray SpectroscopyGainRadiation HardnessSilicon DetectorsSilicon Drift DetectorsSilicon Semiconductor DetectorsSystematic Error AnalysisSystematic Error CalibrationSystematic Error Correction
Abstract The SIDDHARTA-2 experiment at the DAΦNE collider aims to perform the first kaonic deuterium x-ray transitions to the fundamental level measurement, with a systematic error at the level of a few eV. To achieve this challenging goal the experimental apparatus is equipped with 384 Silicon Drift Detectors (SDDs) distributed around its cryogenic gaseous target. The SDDs developed by the SIDDHARTA-2 collaboration are suitable for high precision kaonic atoms spectroscopy, thanks to their high energy and time resolutions combined with their radiation hardness. The energy response of each detector must be calibrated and monitored to keep the systematic error, due to processes such as gain fluctuations, at the level of 2–3 eV. This paper presents the SIDDHARTA-2 calibration method which was optimized during the preliminary phase of the experiment in the real background conditions of the DAΦNE collider, which is a fundamental tool to guarantee the high precision spectroscopic performances of the system over long periods of data taking, as that required for the kaonic deuterium measurement.
20
A new measurement of kaonic hydrogen X-rays
M. Bazzi, G. Beer, L. Bombelli et al. · Physics Letters B · 2011 · 383 citations · Full text
In-medium nuclear interactions of low-energy hadrons
Physics Reports · 2007 · 319 citations · Full text