Review of Scientific Instruments · 2022 · 19 citations · 38 references
EngineeringElectrical EngineeringInstrumentationMicrowave EngineeringOpticsDark MatterDetector TechnologyMicrowave AmplifiersMicrowave MeasurementsMicrowave Noise AnalysisMicrowave Amplifier CharacterizationMicrowave DiagnosticsMicrowave Frequency MetrologyMicrowave Noise CharacterizationNoise TemperatureRf Detectors
In this paper, we will describe the characterization of an RF amplification chain based on a traveling wave parametric amplifier. The detection chain is meant to be used for dark matter axion searches, and thus, it is coupled to a high Q microwave resonant cavity. A system noise temperature Tsys = (3.3 ± 0.1) K is measured at a frequency of 10.77 GHz, using a novel calibration scheme, allowing for measurement of Tsys exactly at the cavity output port.
38
Journal of the Franklin Institute · 1963 · 5.8K citations
Steven Weinberg · Physical Review Letters · 1978 · 5.2K citations
Cosmology of the invisible axion
John Preskill, Mark B. Wise, Frank Wilczek · Physics Letters B · 1983 · 3.3K citations