Biomedical Optics Express · 2022 · 16 citations · 43 references
Natural SciencesCell BiologyOphthalmologyBiomedical OpticsBiophotonicsOcular ImagingCellular ImagingCorneal AnatomyCorneal ImagingOphthalmic ImagingOptical Coherence TomographyTissue OpticsAxial ResolutionCorneal DiagnosticsCorneal Endothelial Cell ImagingCorneal HistologyCorneal Optical Coherence TomographyLateral ResolutionOptical Coherence MicroscopyOptical Coherence Tomography Instrumentation
Corneal imaging is important for the diagnostic and therapeutic evaluation of many eye diseases. Optical coherence tomography (OCT) is extensively used in ocular imaging due to its non-invasive and high-resolution volumetric imaging characteristics. Optical coherence microscopy (OCM) is a technical variation of OCT that can image the cornea with cellular resolution. Here, we demonstrate a blue-light OCM as a low-cost and easily reproducible system to visualize corneal cellular structures such as epithelial cells, endothelial cells, keratocytes, and collagen bundles within stromal lamellae. Our blue-light OCM system achieved an axial resolution of 12 µm in tissue over a 1.2 mm imaging depth, and a lateral resolution of 1.6 µm over a field of view of 750 µm × 750 µm.
43
Submicrometer axial resolution optical coherence tomography
Boris Považay, Kostadinka Bizheva, Angelika Unterhuber et al. · Optics Letters · 2002 · 478 citations