Optics Express · 2003 · 1.8K citations · 7 references
EngineeringBiomedical EngineeringOphthalmologyBiomedical OpticsOpticsPhotonicsFourier OpticsOptical CoherenceOptical Coherence TomographyOptical DiagnosticsOptical ImagingOptical TomographyFourier-domain Optical Coherence TomographyFourier-domain Signal ProcessingLight LevelsNoise SourcesNoise Sources In DetectorsOptical Coherence Tomography InstrumentationTime-domain Optical Coherence Tomography
In this article we present a detailed discussion of noise sources in Fourier Domain Optical Coherence Tomography (FDOCT) setups. The performance of FDOCT with charge coupled device (CCD) cameras is compared to current standard time domain OCT systems. We describe how to measure sensitivity in the case of FDOCT and confirm the theoretically obtained values. It is shown that FDOCT systems have a large sensitivity advantage and allow for sensitivities well above 80dB, even in situations with low light levels and high speed detection.
7
In vivo video rate optical coherence tomography
Andrew M. Rollins, Manish D. Kulkarni, Siavash Yazdanfar et al. · Optics Express · 1998 · 416 citations · Full text