Precision spinning drop interfacial tensiometer

Y. Seeto, L. E. Scriven

Review of Scientific Instruments · 1982 · 15 citations · 1 references

Concepts

Abstract

An improved instrument is described for precise measurement of interfacial tension by means of a spinning drop in gyrostatic equilibrium. Vibrations which can interfere with gyrostatic equilibrium are greatly reduced and bearing heating artifacts are avoided. Sample holders of biprecision tubing permit determination of the drop magnification factor to be replaced by measurement of the liquid refractive index, which is much more precise. Other improvements and cost effective features are described. Systematic errors are analyzed. They increase from 2% as tension drops below 10 mdyn/cm (10−5 N/m), but decrease from 1% for tensions exceeding 0.08 dyn/cm (8×10−5 N/m), in typical circumstances.

References

1