A low volume specimen container suitable for monitoring the aPTT of heparinized patients

M. J. Ray, P. A. Carroll, S. J. E. Just, G. A. T. Hawson

Blood Coagulation & Fibrinolysis · 1993 · 10 citations · 0 references

Concepts

Abstract

Differences in the activated partial thromboplastin time (aPTT) were shown when blood taken from patients receiving intravenous heparin therapy was collected into 5 ml and 1 ml citrate containers. Mean aPTTs were 27% shorter with the plasma from the 1 ml citrate containers (n = 23). These results were paralleled by a 37% reduction in the mean heparin concentration (n = 11) and a 77% increase in the mean platelet factor 4 (PF4) concentration (n = 7). This phenomenon is due to increased platelet activation and subsequent increased heparin neutralization in the 1 ml citrate container. In an attempt to overcome this, the citrate was removed from a 1 ml container and replaced with a buffered tri-sodium citrate solution containing theophylline, adenosine and dipyridamole anticoagulant (CTAD). Blood from heparinized patients taken into both 5 ml citrate and 1 ml CTAD showed a correction of the shortening artefact in the low volume container. The mean aPTT of plasmas from the 1 ml CTAD container showed an increase of 10% compared with the 5 ml citrate. There was no significant difference in the mean heparin or PF4 concentrations of blood taken into either container. The 1 ml CTAD tube described is a suitable collection container for monitoring heparin in neonates or patients who are difficult to venepuncture and overcomes the neutralization of heparin in part filled low volume containers.