Photoacoustics · 2022 · 62 citations · 44 references
EngineeringBioengineeringBiomedical EngineeringBiomedical OpticsBiosensorsSensorsAcoustic SensorsBiomedical SensorsFiber OpticsOptical FibersOptical SensorsSensor MaterialsBiomedical ApplicationsOptical Amplifier DesignPhotoacoustic Imaging InstrumentationPhotoacoustic ProbesPhotoacoustic Sensor DesignPhotoacoustic Transducer Design
A sensor system for exhaled ammonia (NH3) monitoring exploiting quartz-enhanced photoacoustic spectroscopy (QEPAS) was demonstrated. An erbium-doped fiber amplifier (EDFA) with an operating frequency band targeting an NH3 absorption line falling at 1531.68 nm and capable to emit up to 3 W of optical power was employed. A custom T-shaped grooved QTF with prong spacing of 1 mm was designed and realized to allow a proper focusing of the high-power optical beam exiting the EDFA between the prongs. The performance of the realized sensor system was optimized in terms of spectrophone parameters, laser power and modulation current, resulting in a NH3 minimum detectable concentration of 14 ppb at 1 s averaging time, corresponding to a normalized noise equivalent absorption coefficient (NNEA) of 8.15 × 10−9 cm−1 W/√Hz. Continuous measurements of the NH3 level exhaled by 3 healthy volunteers was carried out to demonstrate the potentiality of the developed sensor for breath analysis applications.
44
Increased amount of nitric oxide in exhaled air of asthmatics
Kjell Alving, Eddie Weitzberg, Lundberg Jm · European Respiratory Journal · 1993 · 1.2K citations · Full text
Life Sciences, Immunology, Allergy +13