2021 · 12 citations · 7 references
Medical ElectronicsEngineeringWearable TechnologyBiomedical EngineeringWireless Implantable DeviceBiomedical DevicesImplantable Medical DevicesHuman BodyElectrical EngineeringEnergy HarvestingWireless Power TransmissionImplantable SensorPorcine SkinImplantable DevicesImplantable DevicePower Transfer EfficiencyBiomedical SensorsBioelectronicsElectrophysiologyMedical Devices
This paper presents the design of capacitive wireless charging (CWC) for implantable medical devices using a film-type electrode. We clarify two points using a porcine skin with subcutaneous fat with a high affinity for the human body. One is the operational principle of CWC using a simplified equivalent circuit. The crucial parameters for the design are derived from a theoretical equation on maximum available efficiency, max. The other is to clarify the power transfer efficiency via CWC demonstration. The measured results showed that a max of 47.5% at 308 MHz could be achieved and a of RF-DC could reach about 23% at an input power of 70 mW. This study will enable implantable devices that can be charged during breaks or sleeping at home, and the patient burden will be greatly reduced.
7