IMRT QA and gamma comparisons: The impact of detector geometry, spatial sampling, and delivery technique on gamma comparison sensitivity

J. Steers, Benedick A. Fraass

Medical Physics · 2021 · 19 citations · 47 references

Abstract

Noticeable differences in error sensitivity were observed for different detector geometries, but differences were dependent on induced error type, and a particular device geometry did not offer universal improvements in error sensitivity across studied error types. This study demonstrates that the sensitivity of the gamma comparison does not largely hinge on detector spatial sampling. VMAT deliveries were generally less sensitive to errors when compared to IMRT plans for the planar-type devices, while similar sensitivities were observed between delivery techniques for the ArcCHECK device. Results of this work suggest that a universal gamma criterion is inappropriate for IMRT QA and that the percent pixels passing is an insufficient metric for evaluating quality assurance checks in the clinic.