Medical Physics · 2021 · 19 citations · 47 references
Noticeable differences in error sensitivity were observed for different detector geometries, but differences were dependent on induced error type, and a particular device geometry did not offer universal improvements in error sensitivity across studied error types. This study demonstrates that the sensitivity of the gamma comparison does not largely hinge on detector spatial sampling. VMAT deliveries were generally less sensitive to errors when compared to IMRT plans for the planar-type devices, while similar sensitivities were observed between delivery techniques for the ArcCHECK device. Results of this work suggest that a universal gamma criterion is inappropriate for IMRT QA and that the percent pixels passing is an insufficient metric for evaluating quality assurance checks in the clinic.
47
A technique for the quantitative evaluation of dose distributions
Daniel A. Low, William B. Harms, Sasa Mutic et al. · Medical Physics · 1998 · 2.8K citations · Full text
Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy
Laura Masi, Raffaela Doro, V. Favuzza et al. · Medical Physics · 2013 · 280 citations