Medical Physics · 2021 · 203 citations · 21 references
Silicon PIN photodiode BPW34S, previously tested for X-rays and Co-60 gamma ray source, can also be a reliable candidate for dose rate and absorbed skin dose determination in typical radiotherapy treatments irradiations. A low sensitivity loss below 2% up to 21.4 Gy has been measured, allowing its use as an affordable reusable skin dosimeter. Moreover, no significant difference has been observed between its response to dose-per-pulse and changing pulse repetition frequency in terms of sensitivity and dependence with dose-rate value.
21
Silicon Carbide Detectors for in vivo Dosimetry
G. Bertuccio, Donatella Puglisi, Daniele Macera et al. · IEEE Transactions on Nuclear Science · 2014 · 223 citations · Full text
Electronic dosimetry in radiation therapy
Anatoly Rosenfeld · Radiation Measurements · 2006 · 122 citations