Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery

11

Citations

20

References

2021

Year

Abstract

A commissioning method is recommended using the 3D plate calibration, which is verified by radiation isocenter and validated with couch-angle dependency, baseline drift, and gated-delivered-dose equivalency tests. This method characterizes OSI uncertainties, ensuring motion-monitoring accuracy for SRS treatments.