Sensors · 2021 · 33 citations · 16 references
MedicineBiomedical EngineeringAcousticsMicrofabricationUltrasoundAcoustic ImagingTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound InstrumentationAcoustic Transducer ArraysAxial ResolutionBandwidthImpulse ResponseTransducer FabricationTransducersUltrasonic Transducer ArraysUltrasonic Transducers
Piezoelectric micromachined ultrasonic transducers (PMUT) are promising elements to fabricate a two-dimensional (2D) array with a pitch small enough (approximately half wavelength) to form and receive arbitrary acoustic beams for medical imaging. However, PMUT arrays have so far failed to combine the wide, high-frequency bandwidth needed to achieve a high axial resolution. In this paper, a polydimethylsiloxane (PDMS) backing structure is introduced into the PMUTs to improve the device bandwidth while keeping a sub-wavelength (λ) pitch. We implement this backing on a 16 × 8 array with 75 µm pitch (3λ/4) with a 15 MHz working frequency. Adding the backing nearly doubles the bandwidth to 92% (-6 dB) and has little influence on the impulse response sensitivity. By widening the transducer bandwidth, this backing may enable using PMUT ultrasonic arrays for high-resolution 3D imaging.
16
Yahai Lu, Hao-Yen Tang, Stephanie Fung et al. · Applied Physics Letters · 2015 · 236 citations
Monolithic ultrasound fingerprint sensor
Xiaoyue Jiang, Yipeng Lu, Hao-Yen Tang et al. · Microsystems & Nanoengineering · 2017 · 162 citations · Full text
Engineering, Biomedical Engineering, Microelectromechanical Systems +16