Dose-response of deformable radiochromic dosimeters for spot scanning proton therapy

S. Jensen, Lia Valdetaro, P.R. Poulsen, Péter Balling, Jørgen B. B. Petersen, L.P. Muren

Physics and Imaging in Radiation Oncology · 2020 · 22 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

Intrafractional motion and deformation influence proton therapy delivery for tumours in the thorax, abdomen and pelvis. This study aimed to test the dose-response of a compressively strained three-dimensional silicone-based radiochromic dosimeter during proton beam delivery. The dosimeter was read-out in its relaxed state using optical computed tomography and calibrated for the linear energy transfer, based on Monte Carlo simulations. A three-dimensional gamma analysis showed a 99.3% pass rate for 3%/3 mm and 93.9% for 2%/2 mm, for five superimposed measurements using deformation-including Monte Carlo dose calculations as reference. We conclude that the dosimeter's dose-response is unaffected by deformations.

References