Corrosion of Polished Cobalt-Chrome Stems Presenting as Cobalt Encephalopathy

Stephen S. Tower, Dana J. Medlin, Robert L. Bridges, Christina S. Cho

Arthroplasty Today · 2020 · 12 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

fluid cobalt Cobalt neurotoxicity Quantitative FDGPET brain scan a b s t r a c t Adverse reactions to metallic debris from corrosion of polished cobalt-chromiumecemented femoral stems are reported. Cobaltism (systemic cobalt poisoning) has not been reported from this phenomenon. Three patients presented to their surgeon for ongoing care 10-20 years after primary metal-on-plastic hip arthroplasty with the same polished cobalt-chromiumecemented femoral stems (Heritage, Zimmer). Urine cobalt was elevated, and the patients had symptoms consistent with cobaltism. Quantitative-F 16 DG-PET-CT brain imaging was performed showing generalized and focal brain hypometabolism consistent with cobalt encephalopathy. At revision, all stems were well fixed and grossly corroded. At 1 year after revision, cobalturia and cognitive symptoms were resolved or improved. Mechanically assisted crevice corrosion at the polymethylmethacrylate interface is a complication of polished cobalt-chromium ecemented stems that can result in systemic cobalt exposure and toxic encephalopathy. Our cases had only minor periprosthetic symptoms. Patients implanted with polished cobalt-chromiumecemented stems warrant monitoring with urine cobalt. Patients with cobaltemia warrant an evaluation for toxic encephalopathy.

References

17